Fig.5
Black-Neutral
Red-Phase
Green-Earth
Fig.4
7
11Safety lock button (Only for M-6202)
To select the range for current consumption measurement, turn the range
switch while this button is being pressed. It is to prevent any hazards will
not occur on user by DUT operates suddenly.
12Protective fuse
Protected by fuse of 600V/500mA ceramic fuse (F type Φ6.35mm x 32mm)
against external over voltage.
13Mains cord : M-7125
This mains cord can be connected to the main power supply so that the
instrument can derive power from it. To measure contact current, the socket
of the mains power supply is to be equipped with an earth terminal.
14RPE-Test lead : M-7129
The Test Lead can be combined with either alligator clip or probe with blade
type prod.
Whichever can be used as usage.
15Attachment - Probe with blade type prod
16Attachment - Alligator clip
3.5 Optional accessory
1Test Lead for external terminal: M-7122
It can be used for a measurement via external terminal on the instrument.
2Test Lead for measuring distribution board : M-7121
It can be used when the mains cord (13), described in clause 3.4, cannot
be connected to the outlet (no earthing contact socket is available).
(Please refer to clause 5.2-3 for the connection)