KEW SNAPSeriesKEW SNAP 2413RDIGITAL CLAMP METERKYORITSU ELECTRICAL INSTRUMENTSWORKS,LTD.INSTRUCTION MANUAL
―8―Operating System :Dual IntegrationDisplay :Field effect 3・1/2 digit liquid crystaldisplay with maximum count of 1999Range selection :ManualOverran
① LCDField effect liquid crystal display with maximum indication of 1999.Function symbol(mA, A) and decimal point automatically appear asthe function
―10―“”is displayed on the lower right side for low batterywarning and “1”is displayed only at the highest digit foroverrange indication.② Data Hold Bu
―11―5.Operation5-1 Preparation(1) To Check battery set the function/range switch ⑤ To the desiredposition. If the display is clear without symbol “”sh
―12―CAUTION●Transformers jaws , especially their tips, have been preciselyadjusted to obtain maximum accuracy. Take sufficient care toavoid shock, vib
―13―Note:When measuring large current, observe the time limit specified insection 3, Specifications. Otherwise, the transformers jaws mayoverheat, res
―14―5-3 How to Use Peak Hold Function10ms or 100ms response time can be selected for peak holdmeasurement. Make selection according to your applicatio
―15―Note 2:If it is necessary to read the display away from the conductorin a peak hold measurement, press the data hold switch ②first and then remove
―16―5-4 How to Use The Frequency Selector SwitchKEW SNAP 2413R has a very good frequency response because of theelectromagnetic property of its transf
―17―5-5 How to Use Data Hold FunctionPush in the Data Hold button to freeze the reading. This is especiallyuseful for taking a reading in a dimly lit
Contents1.Safety Warnings ………………………………………………………12.Features ………………………………………………………………43.Specifications …………………………………………………………54.Instrument Layout
―18―KEW SNAP 2413R5-7 Analogue Output: How to Use Model 7073 Output CordAC and DC output can be obtained by inserting optional Model 7073output cord i
Replace the battery when“”symbol appears on the LCD.(1) Set the function/range switch to the OFF position.(2) Unscrew and remove the battery compartme
―20―Use a damp cloth detergent for cleaning the body of the instrument. Toavoid possible deforming or discoloring, do not use solutions containingsolv
―21―Use the following troubleshooting guide for hints on problems withinstrument operation.8.Before Sending for RepairTransformer jaws buzzwhen measur
06-12DISTRIBUTOR92-1870AKyoritsu reserves the rights to change specifications ordesigns described in this manual without notice and withoutobligations
―1―1.Safety WarningsMake sure to read through this instruction manual before using thisinstrument.●The instrument has been designed, manufactured, and
Following symbols are used on the instrument and in the instructionmanual. Attention should be paid to each symbol to ensure your safety. Refer to the
―3―CAUTION●Make sure that the function selector switch is set to anappropriate position before making measurement.●Be sure to set the function selecto
―4―KEW SNAP 2413R is a unique digital clamp meter for both very lowcurrent and high current measurements. Its shielded transformers jawsminimizes the
―5―Measuring Ranges and Accuracy●AC Current 3.Specifications● Effective Value (RMS)Most alternating currents and voltages are expressed in effectiveva
―7―Range200mA2A20A200A1000AMeasuring Range0∼200mA0∼2A0∼20A0∼200A0∼500A501A∼1000AAC Output VoltageAccuracy0∼200mV±2.0%rdg±0.5mV±2.5%rdg±0.5mV±3.0%rdg±0